<APPENDIX 3. OPERATION OF EACH PARAMETER IN FAILURE MODE>

A3-1

IM 01F06F00-01EN

APPENDIX 3. OPERATION OF EACH

PARAMETER IN FAILURE MODE

*1: Standard Type and Multi-variable Type with THERMOMETER_FUNCTION in TR block Set to “Monitor Only” or “Not Use”
*2: Multi-variable Type with THERMOMETER_FUNCTION Used for Density Calculation

LCD

Display

Alarm Detail

RS Block

TR Block

AI1 Block

AI2 Block

AI3 Block

Alarm Reset

SW* (default)

AL-01

AMP. Module
Failure 1 (AL-01)

<BLOCK_ERR>
Other

<PV.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

Not provided

<XD_ERROR>
AMP. Module
Failure 1 (AL-01)
<PV.Status>
Bad-Device
Failure

<OUT.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

<SV.Status>
Bad-Device
Failure

AL-02

COM. Circuit
Failure 1 (AL-02)

<BLOCK_ERR>
Other

<PV.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

Not provided

<XD_ERROR>
COM. Circuit
Failure 1 (AL-02)
<PV.Status>
Bad-Device
Failure

<OUT.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

<SV.Status>
Bad-Device
Failure

AL-03

COM. Circuit
Failure 2 (AL-03)

<BLOCK_ERR>
Other

<PV.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

Not provided

<XD_ERROR>
COM. Circuit
Failure 2 (AL-03)
<PV.Status>
Bad-Device
Failure

<OUT.Status>
• Default
Bad-Non Specifi c
• STATUS_OPTS:Propagate Fault Forward = Active
Bad-Device Failure

<SV.Status>
Bad-Device
Failure

AL-04

AMP. Module
Failure 2 (AL-04)

<BLOCK_ERR>
Lost Static Data
Lost MV Data

<BLOCK_ERR>
Other

<PV.Status>
Bad-Non Specifi c

Not provided

<XD_ERROR>
AMP. Module
Failure 2 (AL-04)
<PV.Status>
Bad-Non Specifi c <OUT.Status>

Bad-Non Specifi c

<SV.Status>
Bad-Non Specifi c

AL-05

Flow Sensor
Failure (AL-05)

<BLOCK_ERR>
Other

<PV.Status>
Uncertain-Non
Specifi c

<PV.Status>
Uncertain-Non
Specifi c

Provided

(ON)

<XD_ERROR>
Flow Sensor
Failure (AL-05)
<PV.Status>
Uncertain-
Sensor
Conversion not
Accurate

<OUT.Status>
Uncertain-Non
Specifi c

<OUT.Status>
Uncertain-Non
Specifi c