Yokogawa digitalYEWFLO (DY-FF) User Manual

Page 40

Advertising
background image

<7. IN-PROCESS OPERATION>

7-2

IM 01F06F00-01EN

LCD

Error Detail

Alarm Mask

SW (default)

AL-86

AR Range High (AR.RANGE_HI) is smaller
than AR Range Low (AR.RANGE_LOW).

Provided (OFF)

AL-87

AR Input1 (AR.IN_1) is over range.

Provided (OFF)

AL-88

AR Input2 (AR.IN_2) is over range.

Provided (OFF)

AL-89

AR Input (AR.IN) is not connected to the
volumetric fl ow.

Provided (OFF)

AL-90

AR Input1 (AR.IN_1) is not connected to the
temperature.

Provided (OFF)

AL-91

AR Input2 (AR.IN_2) is not connected to the
pressure.

Provided (OFF)

AL-92

AR Compensation Coeffi cient (AR.
AR_FLOW_CONFIG.Element) changed
unexpected.
Therefore AR Output (AR.OUT.Value) is
uncertainty.

Provided (OFF)

AL-93

AR Output Range .Units Index (AR.OUT_
RANGE.Unit Index) is not selected rightly the
corresponding to AR Arithmetic Type (AR.
ARITH_TYPE).

Provided (OFF)

*: Not provided for a model with the option /MV and with the fl uid

density calculation set to be active.

LCD

Error Detail

Alarm Mask

SW (default)

AL-01

The EEPROM(S) failed.

Not provided

AL-02

The serial communication circuit in the
amplifi er failed (type 1 error).

Not provided

AL-03

The serial communication circuit in the
amplifi er failed (type 2 error).

Not provided

AL-04

The EEPROM(F) failed.

Not provided

AL-05

The fl ow sensor failed.

Provided (ON)

AL-06

The input circuit in the amplifi er failed.

Provided (ON) *

AL-07

The temperature circuit in the amplifi er failed. Not provided

AL-08

The temperature sensor failed.

Not provided

AL-20

No function blocks are scheduled.

Not provided

AL-21

Resource Block is in O/S mode.

Not provided

AL-22

Transducer Block is in O/S mode.

Not provided

AL-23

AI1 Block is in O/S mode.

Provided (ON)

AL-24

AI2 Block is in O/S mode.

Provided (OFF)

AL-25

DI1 Block is in O/S mode.

Provided (OFF)

AL-26

DI2 Block is in O/S mode.

Provided (OFF)

AL-27

PID Block is in O/S mode.

Provided (OFF)

AL-28

AI3 Block is in O/S mode.

Provided (OFF)

AL-29

IT Block is in O/S mode.

Provided (OFF)

AL-30

AR Block is in O/S mode.

Provided (OFF)

AL-41

Flow rate is over the range.

Not provided

AL-42

The fl ow rate span setting exceeds the range
limit.

Not provided

AL-43

Temperature is over the range. (Regulated in
the upper or lower limit value)

Not provided

AL-51

The transient vibration makes the current fl ow
rate output constant.

Provided (OFF)

AL-52

The high vibration makes the current fl ow rate
output zero.

Provided (OFF)

AL-53

The shedder bar is clogged with a material.

Provided (OFF)

AL-54

The current fl ow rate is fl uctuating more than
20%.

Provided (OFF)

AL-61

Indicator is over the range.

Not provided

AL-62

AI1 Block is in Manual mode.

Provided (ON)

AL-63

AI1 Block is in simulation mode.

Provided (ON)

AL-64

AI1 Block is not scheduled.

Provided (ON)

AL-65

AI2 Block is in Manual mode.

Provided (OFF)

AL-66

AI2 Block is in simulation mode.

Provided (OFF)

AL-67

AI2 Block is not scheduled.

Provided (OFF)

AL-68

DI1 Block is in Manual mode.

Provided (OFF)

AL-69

DI1 Block is in simulation mode.

Provided (OFF)

AL-70

DI1 Block is not scheduled.

Provided (OFF)

AL-71

DI2 Block is in Manual mode.

Provided (OFF)

AL-72

DI2 Block is in simulation mode.

Provided (OFF)

AL-73

DI2 Block is not scheduled.

Provided (OFF)

AL-74

PID Block is in Bypass mode.

Provided (OFF)

AL-75

PID Block is failed (type 1 error).

Provided (OFF)

AL-76

PID Block is failed (type 2 error).

Provided (OFF)

AL-77

AI3 Block is in Manual mode.

Provided (OFF)

AL-78

AI3 Block is in simulation mode.

Provided (OFF)

AL-79

AI3 Block is not scheduled.

Provided (OFF)

AL-80

IT Block is in Manual mode.

Provided (OFF)

AL-81

IT Block is not scheduled.

Provided (OFF)

AL-82

IT Total backup failed. Last IT Output.Value (IT.
OUT.Value) could not saved.

Provided (OFF)

AL-83

IT Clock Period (IT.CLOCK_PER) is smaller
than IT Period of Execution(IT.EXECUTION_
PERIOD).

Provided (OFF)

AL-84

AR Block is in Manual mode.

Provided (OFF)

AL-85

AR Block is not scheduled.

Provided (OFF)

Table 7.1

Alarm Indications and Alarm Mask Switches

Advertising