2 model and suffix codes, 3 optional specifications, Model and suffi x codes -3 – Yokogawa digitalYEWFLO (DY-FF) User Manual

Page 47: 3 optional, Specifi cations -3, 2 model and suffi x codes, 3 optional specifi cations important

Advertising
background image

<9. GENERAL SPECIFICATIONS>

9-3

IM 01F06F00-01EN

Functional Specifi cations:

Functional specifi cations for Fieldbus communication conform to the standard specifi cations (H1) of F

OUNDATION

fi eldbus.
F

OUNDATION

Fieldbus specifi cations (ITK 5.0.1) grant the interoperability of the fi eld instruments.

Function blocks:

Block name

Number

Execution time

Note

AI 3

29

ms

AI1: Monitors the fow rate and totalized fl ow rate; AI2: Monitors the temperature
for a model with the multi-variable type option; AI3: volumetric fl ow input for mass
fl owrate calculation of AR.

DI

2

25 ms

Flow and temperature limit switches

AR 1

40

ms

Mass

fl ow calculation

IT 1

40

ms

Integrator block integrates a variable as a function of the time or accumulates the
counts

PID

1

40 ms

Applicable when LC1 option is specifi ed

9.2 Model and Suffi x Codes

DY

-F -/

-N

DYA-F/
F: digital communication (F

OUNDATION

Fieldbus protocol)

N: Remote type detector

9.3 Optional

Specifi cations

IMPORTANT

In case of the remote type, select the same specifi cation (code) for both detector and converter.

For options other than below, refer to GS 01F06A00-01EN.

(Note1) For intrinsically safe approval, use the barrier certifi ed by the testing laboratories (BARD-400 is not applicable).

Item Description

Code

Multi-variable Type

Provides a temperature sensor (Pt 1000) built-in the vortex shedder bar, enabling the AI2 function block
to output the process fl uid temperature, and mass fl ow rates to be calculated. (For details, refer to GS
01F06A00-01EN.)

MV

PID Function

Provides a PID control function block.

LC1

Software download
function

Based on F

OUNDATION

Fieldbus Specifi cation (FF-883)

Download class: Class 1

EE

Advertising