Contents of DEVICE_STATUS_3 (Index 1047)

Hexadecimal

Display through DD

Description

0x10000000 No FB scheduled (AL-20)

No function blocks are scheduled.

0x02000000 TB in O/S mode (AL-22)

The transducer block is in O/S mode.

0x01000000 AI1 in O/S mode (AL-23)

The AI1 block is in O/S mode.

0x00800000 AI2 in O/S mode (AL-24)

The AI2 block is in O/S mode.

0x00400000 DI1 in O/S mode (AL-25)

The DI1 block is in O/S mode.

0x00200000 DI2 in O/S mode (AL-26)

The DI2 block is in O/S mode.

0x00100000 PID in O/S mode (AL-27)

The PID block is in O/S mode.

0x00040000 AI1 in MAN mode (AL-62)

The AI1 block is in manual mode.

0x00020000 AI1 Simulation Active (AL-63)

Simulation is enabled in the AI1 block.

0x00010000 AI1 not Scheduled (AL-64)

The AI1 block is not scheduled.

0x00004000 AI2 in Man Mode (AL-65)

The AI2 block is in manual mode.

0x00002000 AI2 Simulate Active (AL-66)

Simulation is enabled in the AI2 block.

0x00001000 AI2 not scheduled (AL-67)

The AI2 block is not scheduled.

0x00000400 DI1 in MAN mode (AL-68)

The DI1 block is in manual mode.

0x00000200 DI1 in simulate active (AL-69)

Simulation is enabled in the DI1 block

0x00000100 DI1 not scheduled (AL-70)

The DI1 block is not scheduled.

0x00000040 DI2 in MAN mode (AL-71)

The DI2 block is in manual mode.

0x00000020 DI2 in simulate active (AL-72)

Simulation is enabled in the DI2 block.

0x00000010 DI2 not scheduled (AL-73)

The DI2 block is not scheduled.

0x00000004 PID in BYPASS mode (AL-74)

The PID block is in BYPASS mode.

0x00000002 PID error 1 (AL-75)

PID block error 1

0x00000001 PID error 2 (AL-76)

PID block error 2

Table 8.4

Contents of DEVICE_STATUS_4 (Index 1048)

Hexadecimal

Display through DD

Description

0x00000100 Indicator over range (AL-61)

Indicator overrange

0x00000080 Flow velocity over range (AL-41)

Flow velocity overrange

0x00000040 Flow rate span exceed limit (AL-42)

The fl ow rate span setting exceeds the range limit.

0x00000020 Temp. over range (AL-43)

Temperature overrange

0x00000008 Transient vibration (AL-51)

Transient excessive vibration (transient disturbance)

0x00000004 High vibration (AL-52)

Excessive vibration

0x00000002 Clogging (AL-53)

Flow anomaly (clogging)

0x00000001 Fluctuating (AL-54)

Flow anomaly (excessive output fl uctuations)