<10. EXPLOSION PROTECTED TYPE INSTRUMENT>

10-3

IM 01F06F00-01EN

Maintenance and Repair

WARNING

• The instrument modifi cation or parts

replacement by other than authorized
representative of Yokogawa Electric
Corporation is prohibited and will void the
certifi cation.

Installation Diagram of Intrinsically safe (and Note)

F1002.ai

DY

(Flowmeter)

Terminator

Terminator

Safety Barriar

Field Instrument

Field Instrument

Hazardous Location

Non Hazardous Location

DYA

(Converter)

DY-N

(Flowmeter)

Terminator

Terminator

Safety Barriar

Field Instrument

Field Instrument

Hazardous Location

Non Hazardous Location

A
B
T(*1)

A
B
T
C

DYC (Signal Cable)

[ Integral type ]

[ Remote type ]

(*1): Wire for T termanal

With temperature sensor type: Installed
Without temperature sensor type: Not Installed

Note

‡

In the rating 1, the output current of the barrier must be limited by a resistor ‘Ra’ such that Io=Uo/Ra.

‡

In the rating 2, the output of the barrier must be the characteristics of the trapezoid or the rectangle and this transmitter can be

connected to Fieldbus equipment which are in according to the FISCO model.

‡

The terminators may be built-in by a barrier.

‡

More than one field instrument may be connected to the power supply line.

‡

The terminator and the safety barrier shall be certified.

Electrical data

II C

II B

Rating1 (Entity)

Rating2 (FISCO)

Rating3 (FISCO)

Maximum Input Voltage Ui

24V

17.5V

17.5V

Maximum Input Current Ii

250mA

380mA

460mA

Maximum Input Power Pi

1.2W

5.32W

5.32W

Maximum Internal Capacitance Ci

1.76nF

1.76nF

1.76nF

Maximum Internal Inductance Li

0

0

0