IM 12D06A03-01E

Appendix 3-11

13-3. ISC202G, ISC202S 2-Wire Inductive Conductivity Transmitter (Fieldbus Communication)

Quality
Inspection
Standards

ISC202G, ISC202S
2-Wire Inductive Conductivity Transmitter
(Fieldbus Communication)

QIS 12D06A03-61E

1st Edition: Apr. 2007

1. Scope

This inspection standard applies to the ISC202G and ISC202S 2-Wire Inductive Conductivity

Transmitters (Fieldbus specification).

2. Inspection

Items

2.1 Insulation resistance test

* 2.2 Dielectric strength test

2.3 Temperature indication check

* 2.4 Conductivity indication check

* 2.5 Fieldbus communication functional check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.

3.

Inspection Methods, Standards and Conditions

x

Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least

5 minutes before conducting the tests. For the connections for the insulation resistance test,

follow the instructions in Sections 3.1and for the connections for the dielectric strength test,

follow the instructions in Section 3.2.

x

Use test equipment shown in Figure 1, or equivalent, for the tests.

3.1 Insulation Resistance Test

x

Apply 500 V DC between the power supply terminals (+ and –) plus input terminals (11 to

17), shorted together, and the earth terminal (G). The insulation resistance must be 100 MΩ

or greater.

3.2 Dielectric Strength Test (Required Only for ISC202S)

Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz

or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must

withstand this voltage. (The sensed current should be 10 mA.)

x

Between the power supply terminals (+ and –) plus input terminals (11 to 17), shorted

together, and the earth terminal (G)

3.3 Temperature

Indication

Check

Connect the instruments as shown in Figure 1, and set them as follows.

Decade resistance box 1: 1097.3 :

Decade resistance box 2: 100 Ω

Set the type of temperature sensor to PT1000.

Setting Service Code 10 to 1.

a. Press the [ ] key.

b. Press the [NO] key twice. The message display will show “ SERV.”

c. Press the [YES] key. The data display will show “00” with the first digit of 0 flashing.

d. Press the [

š

] key once. The data display will show “10.”

e. Press the [ENT] key. The message display will show “ T.SENS.”

f. Press

the

[

š

] key once. The data display will change from “0” to “1.”

g. Press the [ENT] key until the message display shows “ SERV.”

h. Press the [ ] key. The instrument returns to the measurement mode.