2. commissioning mode, 2. commissioning mode -6 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 48

Advertising
background image

IM 12D06A03-01E

5-6 Parameter setting

5-2. Commissioning mode

In order to obtain peak performance from the EXA ISC202, you must set it up for each custom application.

*OUTP

: mA output is set as default to 0-1000 mS/cm.

For enhanced resolution in more stable measuring processes, it may be desirable to

select for example 0-100 μS/cm range.

*HOLD

: The EXA ISC202 transmitter has the ability to ‘HOLD’ the output during maintenance

periods. This parameter should be set up to hold the last measured value, or a fixed

value to suit the process (Section 5-2-2).

*TEMP.1, 2

: First/second temperature compensation types and values (see section 5-2-4 and 5-2-5).

* NaCl is used for neutral salt solutions. Strong solutions of salts are compensated, as

are process waters, pure, and ultrapure water.

* T.C. temperature coefficient compensation uses a linear temperature compensation

factor. This can be set by calibration (section 5) or configuration (service code 21).

* MATRX compensation is an extremely effective way of compensation. Choose from

standard matrix tables, or configure your own to exactly suit your process.

The default setting for *TEMP.1 and *TEMP.2 is NaCl. *TEMP.2 is enabled when it is

not equal to *TEMP.1.

*SERV

: This selection provides access to the service menu.

mA

mA

YES NO

YES NO

YES NO

YES NO

YES NO

OUTPUT

SET HOLD
TEMP.
SERVICE

MODE

MEASURE
CAL

DISPLAY
HOLD

Only if *OUTP. F is set as linear service code 31

Press the COMMISSIONING key

Enables HOLD

Sets Temperature compensation method for

first compensated conductivity

Enables and sets Temperature compensation method for

second compensated conductivity