IM 12D06A03-01E

Appendix 2-1

12. APPENDIX 2

12-1. Preface

This appendix contains these items.

1. Method of wiring and parameter setting the following:.

PH201G*B Dedicated Distributor

BA2O

Junction Terminal Box

WF1OJ Extension

Cable

2. Quick reference for parameter setting

3. Installation factor adjustment.

To ensure that this measurement system can be operated safely and also exhibit its full performance, be

sure to read this appendix before use.

This appendix does not describe SDBT Distributor which is the component unit of the ISC2O2 two-wire

inductive conductivity transmitter system. This unit comes with an instruction manual, so read the instruc-

tion manual IM 1B04T02-01E for details of the unit concerned.

12-2. Wiring diagrams

14

13

Output

(1-5V DC)

A(+)

B(-)

(+)C

(-)D

17

+

-

G

11

12

16

15

Inductive Conductivity

transmitter ISC202G

When PH201G (Style B)

distributor is used

*1

Output

(1-5V DC)

1(+)

2(-)

(+)A

(-)B

SDBT

distributor

*1

Class D (100 ohm or less)

ground

*

2

b

a

d

c

HOLD FAIL

Output

(1-5V DC)

(+)F

(-)H

Relay contact

Non-explosionproof system

Inductive Conductivity

detector ISC40GJ

Output

(1-5V DC)

(+)F

(-)H

*1 Use two-wire cable with OD (Outside Diameter) of 6 to 12 mm

Make sure distributor voltage does not drop below minimum for ISC202G

*2 Ground ISC202G ( Class D ground: 100 ohm or less)

Junction Terminal

Box BA20

Extension Cable

WF10J