2-4 Specifications

NEPSI Certification (ISC202S-K)

NEPSI Intrinsically Safe Type

Cert No. GYJ081158X

• Applicable Standard:

GB3836.1-2000, GB3836.4-2000

• Type of Protection and Marking Code:

Ex ia IIC T4/T6

• Ambient Temperature :

T6; –10 to 40°C, T4; –10 to 55°C

Note 1 Entity Parameters

• Intrinsically safe input parameters

(terminal + and -):

Maximum Input Voltage (Ui) = 31.5 V

Maximum Input Current (Ii) = 100 mA

Maximum Input Power (Pi) = 1.2 W

Maximum Internal Capacitance (Ci) = 22 nF

Maximum Internal Inductance (Li) = 35 μH

• Intrinsically safe output parameters and maximum

external parameters

(terminal 11 and 17):

Uo=14.4 V, Io=20 mA, Po=190 mW, Co=600

nF, Lo=88 mH

Note 2 Installation

• Electrostatic charges on the display window shall

be avoided.

• The external earth connection facility shall be

connected reliably.

• The instrument modification or parts replacement

by other than authorized representative of

Yokogawa Electric Corporation and will void

NEPSI Intrinsically safe certification.

• The user shall not change the configuration in

order to maintain/ensure the explosion protection

performance of the equipment. Any change may

impair safety.

• For installation, use and maintenance of

the product, the end user shall observe the

instruction manual and the following standards:

GB50257-1996 "Code for construction and

acceptance of electric device for explosion

atmospheres and fire hazard electrical

equipment installation engineering''.

GB3836.13-1997 "Electrical apparatus for

explosive gas atmospheres Part 13: Repair and

overhaul for apparatus used in explosive gas

atmospheres".

GB3836.15-2000 "Electrical apparatus for

explosive gas atmospheres- Part 15: Electrical

installations in hazardous area (other than

mines)" .

GB3836.16-2006 "Electrical apparatus for

explosive gas atmospheres- Part 16: lnspection

and maintenance of electrical installation (other

than mines)".

mA-HART® communication
A. Input

: Two wire system 4-20 mA

B. Power supply

:

ISC202G :

up to 40 volts

ISC202S :

up to 31.5 volts

Note:

The transmitter contains a switched

power supply, drawing its energy

from the 0-4 mA section of the

signal. Consequently the 17 volt

limit is applied at 4 mA. The

characteristic of the unit is such that

above about 7 mA on the output, the

terminal voltage can drop to 14.5

volts without problem.

C. Transmission:

Isolated output of 4 to 20 mA DC.

D. Signal

: Maximum load 425Ω at 24 VDC.

(see fi gure 2-1)

Burn to signal failure acc.

NAMUR Recommendation NE43

(18.01.1994)

E. Operating range :

3.9 to 21 mA

F. Communication

:

HART®, 1200 Baud, FSK

modulated on 4 to 20 mA signal

G. Configuration

: Local with 6 keys

H. Software :

Firmware based on Yokogawa stack.

I. Hardware :

Yokogawa HART® Modem F9197UB

J. Other Control systems

:

Yokogawa PRM, Rosemount

AMS, Siemens PDM

K. Hand Terminal :

Rosemount HHT 275/375

L. Other control systems:

Yokogawa PRM, Rose-

mount AMS, Siemens PDM\

M. Output span

:

- Conductivity : min 0.01μS/cm, max. 1999 mS/

cm.

(max 90% zero suppression)

- Resistivity : min 0.001k

Ω

·cm, max.

999 M

Ω

·cm.

(max 90% zero

suppression) The

instrument

is user programmable for

linear or non-linear conductivity

ranges.

N. Cable specification

:

0.5 mm diameter or 24 AWG

over maximum length of 1500 m

O. DD specification

:

The ISC202 Device Description

is available enabling

communications with the

Handheld Communicator and

compatible devices.