3-3. temperature compensation functions, 3-3. temperature compensation functions -14 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 56

Advertising
background image

IM 12D06A03-01E

5-14 Parameter setting

5-3-3. Temperature compensation functions

Code 20

*T.R.°C

Choose a temperature to which the measured conductivity (or resistivity) value

must be compensated. Normally 25°C is used, therefore this temperature is

chosen as default value. Limitations for this setting are: -20 to 140 °C.

If *T.UNIT in code 11 is set to °F, default value is 77°F and the limitations are

0 - 280°F.

Code 21

*T.C.1, 2

In addition to the procedure described in section 5-2-4 and 5-2-5 it is possible

to adjust the compensation factor directly. If the compensation factor of the

sample liquid is known from laboratory experiments or has been previously

determined, it can be introduced here.

Adjust the value between 0.00 to 3.50 % per °C. In combination with reference

temperature setting in code 20 a linear compensation function is obtained,

suitable for all kinds of chemical solutions.

Code 22

*MATRX

The EXA is equipped with a matrix type algorithm for accurate temperature

compensation in various applications. Select the range as close as possible

to the actual temperature/concentration range. The EXA will compensate by

interpolation and extrapolation. Consequently, there is no need for a 100%

coverage.

If 9 is selected the temperature compensation range for the adjustable matrix

must be configured in code 23. Next the specific conductivity values at the

different temperatures must be entered in codes 24 to 28. Matrix data is

exeplified in Appendix 11-3

Code 23 *T1, *T2, *T3, Set the matrix compensation range. It is not necessary to enter equal

*T4 & *T5 °C temperature steps, but the values should increase from T1 to T5, otherwise

the entrance will be refused. Example: 0, 10, 30, 60 and 100 ºC are valid

values for the T1....T5. The minimum span for the range (T5 - T1) is 25 °C. The

valid range for a temperature value is -20º to 140ºC.

Code 24-28 *L1xT1 - In these access codes the specific conductivity values can be entered for

L5xT5

5 different concentrations of the process liquid; each one in one specific access

code (24 to 28). The table below shows a matrix entering example for 0.5 - 5%

H

2

SO

4

solution for a temperature range from 0 - 100 °C. Conductivity range

from 0.0 μS/cm to 1999 mS/cm.

NOTES:

1.

In chapter 11 a table is included to record your programmed values. It will make programming

easy for duplicate systems or in case of data loss.

2.

Each matrix column has to increase in conductivity value.

3.

Error code E4 occurs when two standard solutions have identical conductivity values at the

same temperature within the temperature range.

Table 5-2. Default of user adjustable matrix

T1

T2

T3

T4

T5

Code 23

Temp.

0 °C

25 °C

50 °C

75 °C

100 °C

Code 24

Solution 1

L1

33.8 mS/cm

47.0 mS/cm

57.5 mS/cm

63.7 mS/cm

68.0 mS/cm

Code 25

Solution 2

L2

63.5 mS/cm

92.3 mS/cm

112.5 mS/cm

126.0 mS/cm

137.5 mS/cm

Code 26

Solution 3

L3

95.0 mS/cm

135.3 mS/cm

166.0 mS/cm

188.5 mS/cm

206.0 mS/cm

Code 27

Solution 4

L4

124.5 mS/cm

178.0 mS/cm

220.0 mS/cm

249.0 mS/cm

273.0 mS/cm

Code 28

Solution 5

L5

154.0 mS/cm

218.0 mS/cm

270.0 mS/cm

307.0 mS/cm

336.0 mS/cm

Advertising