IM 12D06A03-01E

Parameter setting 5-13

5-3. Service

Codes

Don't set or input service code numbers other than the code numbers defined in this manual. Setting an

undefined service code may make the transmitter malfunction.

When an undefined service code is input by some accident, push the MODE key and escape from the

service level.

5-3-1. Parameter specific functions

Code 03

*C.C.

Enter the factory calibrated cellconstant mentioned on the textplate or on the

fixed cable. This avoids the need for calibration. Any value between 0.2 and

19.99/cm may be entered.

Code 04

*AIR

To avoid cable influences on the measurement, a ‘zero’ calibration with a

dry sensor may be done. If a connection box (BA10) and extension cable

(WF10) are be used, ‘zero’ calibration should be done including this connection

equipment.

5-3-2. Temperature measuring functions

Code 10 *T.SENS

Selection of the temperature compensation sensor. The default selection is the

30k NTC sensor, which gives excellent precision with the two wire connections

used. The other option gives the flexibility to use a very wide range of other

conductivity/inductive sensors.

Code 11 *T.UNIT

Celsius or Fahrenheit temperature scales can be selected to suit user

preference.

Code 12 *T.ADJ

With the process temperature sensor at a stable known temperature, the

temperature reading is adjusted in the main display to correspond. The

calibration is a zero adjustment to allow for the cable resistance, which will

obviously vary with length.The normal method is to immerse the sensor in a

vessel with water in it, measure the temperature with an accurate thermometer,

and adjust the reading for agreement.

Code Display

Function

Function

detail

X Y

Z Default

values

Parameter specific functions
03

*C.C.

Set cell constant

Use

keys to set value

1.88

/cm

04

*AIR

Zero calibration

Zero calibration with dry cell connected

*START

Press YES to confirm selection

*‘WAIT’

Press YES to start, after briefly displaying

*END

‘WAIT’, *END will be displayed

Press YES to return to commissioning mode

Code Display

Function

Function

detail

X Y Z Default

values

Temperature measuring functions
10

*T.SENS

Temperature sensor 30k NTC

0

0 30k NTC

Pt1000

1

11

*T.UNIT

Display in °C or °F

°C

0

0

°C

°F

1

12

*T.ADJ

Calibrate temperature Adjust reading to allow for cable resistance.

0.0 ºC

Set value relative to current temperature

Use keys to adjust value