Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 83

Advertising
background image

IM 12D06A03-01E

Appendix 2-11

Sample 1
NaCl concentration

Meas. range

250-320 g/l

Reference temp.

60°C

Temp. compensation

Use NaCl compensation (default)

Concentration

― conductivity relationship is linear

NaCl

OUTP

TEMP

60°C

START

Select 1

0% out --> "25%" representing 250 g/l

100% out --> "32%" representing 320 g/l

(g/l display not supported, so here using % instead. Also

display is XX.X%, so we display actual concentration

divided by 10

Calibrate conductivity value using personal SC meter

(1) Set temp. compensation of both ISC202 and personal

SC meter to zero.

(2) Place both ISC202 sensor and personal SC meter sensor

in measured solution, wait for temperature and conductivity

readings to stabilize.

(3) When stabilize, calibrate by adjusting ISC202 reading to same as

personal SC meter reading.

(4) Finally revert temperature compensation to NaCl default.

Service Code 20

0% out -> 404 mS/cm

100% out -> 450 mS/cm

(These conductivity values apply to reference temperature)

Service Code 55

Calibration in Air

Calibration

Sample1E.eps