2-2 Specifications

0

200

400

600

800

12

16

17

20

24

28

32

425

775

Voltage (V)

Load Resistance (

Ω

)

31.5 V

Possible

Fig.2-2 Supply voltage/ load diagram for the ISC202S

J) Temperature compensation:

- Sensor types: Pt1000Ω or 30kΩ NTC

- Automatic: -20 to 140 ºC (0 to 280 ºF)

- Algorithm: selectable as mentioned below

NaCl according to IEC 60746-3

tables.

Two T.C. setting possible between

0.00 to 3.50 %/°C

Matrix: user selectable/

configurable. 8 selectable for

concentrated solutions, 1 free

programmable.

K) Sensor diagnostics:

Abnormal temperature (open, short), abnormal

conductivity values (E5/E6 free programmable),

e.g. dry cell, wiring problems.

L) Calibration:

Manual, calibration Input pre-measureds data

(cell constant).

M) Logbook:

Software record of important events and diag-

nostic data.

N) Display:

Custom liquid crystal display.

- Main display: 3

½

digits, 12.5 mm high, zero

change included.

- Message display: 6 alphanumeric

characters, 7 mm high.

- Special fields: Flags for status indication :

FAIL and HOLD.

- Measuring units: μS/cm or mS/cm

- Key prompts: YES, NO, >, ^, ENT, Menu pointer

- Keys:

6 keys operated through

flexible window with tactile

feedback. One hidden key

behind the front cover.

O) Power supply:

Power supply : Normal 24 V DC loop powered

system, see Figure 2-1, 2-2.

ISC202G: 17 - 40 V DC

ISC202S: 17 - 31.5 V DC

- Input Isolation: Maximum 1000 VDC

mA

mA

P) Housing

:

- Material :

Cast aluminium case with

chemically resistant coat-

ing, cover with flexible

polycarbonate window.

- Color : Case :

Off-white (Equivalent to

Munsell 2.5Y8.4/1.2)

Cover :

Deepsea Moss

green(Equivalent to Mun-

sell 0.6GY3.1/ 2.0)

- Cable gland :

2-Pg13.5

Q) Mounting

:

Pipe, Wall or Panel.

R) Shipping details

:

Package size : W x H x D, 290 x 300 x 290 mm

(11.5 x 11.8 x 11.5 inch).

S) Environment and operational conditions:

- Ambient temp.:

-10 to 55 ºC (+10 to +130 ºF)

LCD operational

temperature is specified -

10 to 70 ºC (14 to 160 ºF)

Excursions to -30 to +70 ºC

will not damage the instrument.

- Storage temp.:

-30 to +70 ºC (-20 to +160 ºF).

- Relative humidity: 10 to 90% RH at 40 ºC

ambient temperature, non

condensing

- Data protection:

EEPROM for configuration

and logbook. Battery

supported clock.

- Watchdog timer : Checks microprocessor.

- Automatic safeguard: Return to measurement

after 10 minutes when

no keystroke.

Operation

protection:

3 digital pass codes (pro-

grammable).

Power

down:

No effect, reset to meas-

urement.

T) HART® specifications

:

Minimum cable diameter: 0.51 mm, 24 AWG.

Maximum cable length: 1500 m

Refer to standard HART® specifications for more

details.

See www.hartcomm.org

U) EMC Conformity standard

,

EN 61326-1 Class A, Table 2

(For use in industrial locations)

EN

61326-2-3

EN 61326-2-5 (pending)

CAUTION

This instrument is a Class A product, and it is

designed for use in the industrial environment.

Please use this instrument in the industrial

environment only.

mA

mA