4. configuration checklist for isc202, 5. coded service settings (default), 4. configuration checklist for isc202 -3 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 71: 5. coded service settings (default) -3

Advertising
background image

IM 12D06A03-01E

Appendix 1-3

Primary choices

Defaults

Alternatives

Reference Menu or

on page

Service code (SC)

Measurement Conductivity

1st Temp. compensation

NaCl in water

Fixed T.C., Matrix

5-2-4

‘Commissioning’

2nd Temp. compensation

NaCl in water

Fixed T.C., Matrix (inactive)

5-2-5

‘Commissioning’

2nd Line display

Process temp.

1st compensation method, Weight%,

5-1-3/4

‘Maintenance’

Output (mA), C.C., Ref. Temp., Software

release, 2nd compensated Conductivity,

2nd compensation method

Range (Linear)

0-1000 mS/cm

0-1999 mS/cm

5-2-1

‘Commissioning’

Temperature compensator

30k NTC

Pt 1000

5-3-2

SC 10

Temperature unit Sensor

Celsius (ºC)

Fahrenheit ( ºF )

5-3-2

SC 11

Cell constant

1.88 /cm

Any value between 0.2 and 19.99 /cm

5-3-1

SC 03

Communication

HART® enabled

disable HART®, PH201*B

5-7

SC 60

Burn out

inactive

HI or LO, Pulse burnout

5-5

SC 32

HOLD during maintenance inactive

Hold last value or fixed value

5-2-2

‘Commissioning’

Calibration temperature

inactive

Adjustment +/- 15ºC

5-3-2

SC 12

AIR (zero) calibration

inactive

Adjustment +/- 15 μS/cm 5-3-1

SC

04

C.C. Calibration

inactive

0.2 and 19.99 /cm

5-1-2

‘Maintenance’

Diagnostics

Hard alarm (all errors) Hard or soft choices

5-6

SC 53

Instrument Limit: E5 (high)

3S

E5 (High) 0.10 - 9.99 S

5-6

SC 54

Instrument Limit: E6 (low)

0.5 μS

E6 (Low) 0.00 - 99.9 μS 5.6

SC

54

Password protection

inactive

(In)active password for different levels

5-6

SC 52

Output in Concentration units inactive

Linearization of output, Weight% on LCD 5-6

SC 55

Code Display

Default Values

Setting #1

Setting #2

Setting #3

03 *C.C. 1.88

/cm

10

*T.SENS

0 : 30 kNTC

11

*T.UNIT

0 : ºC

12 *T.ADJ 0

ºC

20 *T.R.

ºC 25ºC

21

*T.C.1

2.1% per ºC

*T.C.2 2.1%

per

ºC

22

*MATRX

1 : H

2

SO

4

31 *OUTP.F 0:

Linear

32

*BURN

0: No Burnout

50 *RET 1:

On

52 *PASS 0.0.0.:

off.off.off

53

*Err. 05

1 : HARD

*Err. 06

1 : HARD

*Err. 07

1 : HARD

*Err. 08

1 : HARD

54 *E5.L 3S
*E6.LIM

5μS

*OUTP

0% : 0μS

100% : 1000mS

*HOLD

disabled

*TEMP.1

NaCl

*TEMP.2

NaCl

60

*COMM.

1.0: On, write enable

mA

mA

mA

mA

mA

mA

mA

mA

11-5. Coded service settings (default)