6. One-point Calibration example
Normally you would do a laboratory analysis when the sensor is installed and use the resulting value to
calibrate the transmitter. If the concentration at start-up time is known, then you can use concentration
to conductivity tables to determine the conductivity, and use that.
For example, suppose that the concentration is known to be 34%, and the transmitter should display the
corresponding conductivity which is 772mS/cm:
.
CALIB
START
772 mS/cm
CAL.END
ENT
YES
YES
YES
MODE
example6E.eps