2-18 Appendix
6. One-point Calibration example

Normally you would do a laboratory analysis when the sensor is installed and use the resulting value to

calibrate the transmitter. If the concentration at start-up time is known, then you can use concentration

to conductivity tables to determine the conductivity, and use that.

For example, suppose that the concentration is known to be 34%, and the transmitter should display the

corresponding conductivity which is 772mS/cm:

.

CALIB

START

772 mS/cm

CAL.END

ENT

YES

YES

YES

MODE

example6E.eps