IM 12D06A03-01E

Appendix 2-13

Sample 3
Concentration of sulfuric acid mixture

Meas. range

93-97%

Reference temp.

50°C

Temp. compensation

Enter temperature compensation data into a 5 x 5 matrix

sample3E.eps

START

Service code 20

Service code 31

Service code 35

50°C(Default)

Select 1

0% output -->93%

100%output -->97%

Service code 55

TEMP

Calibration in Air

Calibration

NaCl -->NO

TC -->YES

( here use [MODE] key to escape )

Service code 22

Matrix code 9

93
94
95
96
97

25

133
128
122

110

95

37.5

177
171
160
146
125

50

227
217
204
186
162

62.5

283
270
253
230
196

75

344
328
306
288
238

Service code 23

25 to 75°C

Service code 24 to 28

Conductivity unit (mS/cm)

Enter conductivity values corresponding to output current values.

Existing table (below) relates output current to conductivity at 50°C

Output Current (%)

0

25
50
75

100

H

2

SO

4

conc (%)

93
94
95
96
97

Conductivity (mS/cm)

227
217
204
186
162

(%)

°C

Calibrate conductivity value in actual operation
e.g. 95% H

2

SO

4

corresponds to conductivity 204 mS/cm (see above table)

Note: This application requires special sensor body material