Operation; display functions and setting, 1. operator interface, Operation; display functions and setting -1 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 39: 1. operator interface -1

Advertising
background image

IM 12D06A03-01E

Operation 4-1

4. OPERATION; DISPLAY FUNCTIONS AND SETTING

4-1. Operator interface

This section provides an overview of the operation of the EXA operator interface. The basic procedures

for obtaining access to the three levels of operation are described briefly. For a step-by-step guide to

data entry, refer to the relevant section of this user’s manual. Figure 4-1 shows the EXA operator inter-

face.

LEVEL 1: Maintenance

These functions are accessible by pushbutton through a flexible front cover window. The functions make

up the normal day-to-day operations that an operator may be required to complete. Adjustment of the

display and routine calibration are among the features accessible in this way. (See table 4-1).

LEVEL 2: Commissioning

A second menu is exposed when the EXA front cover is removed and the display board is revealed.

Users gain access to this menu by pressing the button marked

*

in the lower right of the display board.

This menu is used to set such values as the output ranges and hold features. It also gives access to the

service menu. (See table 4-1).

LEVEL 3: Service

For more advanced configuration selections, press the button marked

*

, then press ‘NO’ repeatedly

until you reach SERVICE. Now push the ‘YES’ button. Selecting and entering ‘Service Code’ numbers in

the commissioning menu provide access to the more advanced functions. An explanation of the Service

Codes is listed in chapter 5 and an overview table is shown in chapter 11.

Measurement Mode

MODE key

* key

CALIB(ration)

NO key

*OUTP Output Range Setting

NO key

DISP.1

NO key

*HOLD HOLD Settting

NO key

DISP.2

NO key

*TEMP.1 Temperature Setting

NO key

HOLD

NO key

· Select desired mode and press YES

· The MODE key is used as a "Cancel and Return to Measurement Mode" escape key

*TEMP.2 Temperature Setting

NO key

*SERV Service Mode

NO key

Maintenande Mode

Commissioning Mode

Service Mode

Can operate with front panel shut

Need to open front panel cover to operate

Table 4-1. Operations overview

Routine Function

Section

Maintena nce

CALIB

DISP.1, 2

HOLD

Calibration with a standard solution or sample

Read auxiliary data or set message display

Switch hold on/off (when activated)

6

4, 5

5

Commissioning

*OUTP

*HOLD

*TEMP.1, 2

Adjust the output range

Activate the hold function

Select method of temperature compensation

5

5

5

Service
(Access to coded entries from

the commissioning level)

*SERV

Fine tune the specialized functions of the

transmitter

5

NOTE:

All three levels may be separately protected by a password.

See Service Code 52 in chapter 5 Service Code table for details on setting passwords.

mA

mA

mA

mA

Advertising