2. calibration procedure, 2. calibration procedure -2 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 64

Advertising
background image

IM 12D06A03-01E

6-2 Calibration

ENT

NO

YES

MODE

MEASURE
CAL

DISPLAY
HOLD

MODE

6-2. Calibration procedure

Press the MODE key

The

legend

CALIB

appears,

and

the

YES/NO

key prompt flags flash





Put the sensor in standard solution

Press YES




Set the value using the

key

Select the flashing digit with the

key

Increase it’s value by pressing the

key

When the correct value is displayed,

press

ENT

to enter the change


After briefly displaying WAIT,

the

CAL.END

message appears

The calibration is now complete

Put the sensor back in the process and press YES

The cell constant is automatically updated after the calibration and the new value can be read on the dis-

lay as described in section 5-1-2. and 5-1-3.

If the calibrated cell constant is not within the range of 0.2 - 19.99 cm

-1

, error E3 is displayed.