IM 12D06A03-01E

Parameter setting 5-17

Code Display

Function

Function

detail

X

Y

Z

Default

values

mA Outputs
30

Not

used

31

*OUTP.F

mA output functions

Linear

0

0 Linear

Table

1

32

*BURN

Burn function

No burnout

0

0 No Burn.

Burnout

downscale

1

Burnout

upscale

2

Pulse

burnout

3

33, 34

Not used

35

*TABLE

Output table for mA

*0%

Linearisation table for mA in 5% steps.

*5%

The measured value is set in the main

*10%

display using the

keys, for

...

each of the 5% interval steps.

...

Where a value is not known, that value may

*95%

be skipped, and a linear interpolation will

*100%

take

place.

36-49

Not

used

mA

Concentration Output function is done in de following order:

• Set *OUTP.F. (Service Code 31) to table

• Set the Concentration range in % (Service Code 55)

• Set table values (%output and Conductivity values) in *TABLE (Service Code 35)