Troubleshooting, 1. introduction, 2. self diagnostics of the conductivity sensor – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 66: 3. self diagnostics of the temperature sensor, 4. self diagnostics of the electronics, 5. checking during operation, Troubleshooting -1, 1. introduction -1, 2. self diagnostics of the conductivity sensor -1, 3. self diagnostics of the temperature sensor -1

Advertising
background image

IM 12D06A03-01E

Trouble shooting 8-1

8. TROUBLESHOOTING

8-1. Introduction

The EXA ISC202 microprocessor based conductivity analyser continuously monitors the condition of

all key components of the measuring system to ensure that measurement is dependable. If a fault is

detected this is immediatly signalled. Errors are shown on the display with a code. Table shows the

errors which can be detected and gives information to help locate the fault or identify the error. Faults

detected while the instrument is on line can also be signalled by a burnout (section 5-5).

8-2. Self diagnostics of the conductivity sensor

During measurement the instrument adjusts the measuring parameters to give the best conditions for

the actual value being measured. At all values the instrument checks the signal from the cell to search

for distortion. If there is a problem with the installation of the cell and this becomes defective this will

trigger an error message on the display possibly accompanied by a by a burnout signal (section 5-5).

8-3. Self diagnostics of the temperature sensor

The temperature sensor, which is normally built into the conductivity cell, is checked to detect damage

or faulty connections.

8-4. Self diagnostics of the electronics

The microprosessor opertion is checked by a watchdog which initiates an electronic reset if the normal

functions suffers severe interference. During reset the instrument checks the program and all stored

data.

If a fault is then detected an alarm is given.

8-5. Checking during operation

Whenever the instrument is being programmed or calibrated, data is checked and an error is shown

when appropriate. Should this occur the new data is rejected and the instrument continues to work with

the previuos settings.

Advertising