3. peripheral products, 3-1. ph201g*b dedicated distributor – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 74

Advertising
background image

IM 12D06A03-01E

2-2 Appendix
12-3. Peripheral products
12-3-1. PH201G*B Dedicated Distributor
Model and suffix codes

Model

Suffix Code

Option Code

Description

Distributor
100V AC
220V AC
Style B
Terminal for Power connection

/TB

-A1
-A2

*B

PH201G
Power Supply

Option

T2.2.2E.EPS

Communication setup with PH201G (style B) distributor (service code 60)

This communication is a one-way to PH201G (Style B) distributor, a power supplier for the EXA 202

transmitters. The PH201G (Style B) receives a current signal (4-20 mA DC) and a digital signal superim-

posed on the DC signal. In other words, the PH201G (Style B) provides a measurement signal, a hold-

contact signal and a fail-contact signal. The communication with PH201G (Style B) is set in Service code

60.

Code 60 *COM The settings should be adjusted to suit the communicating device connected to the out-

put. The communication can be set to HART® or to PH201G (Style B) distributor.

When used with our PH201G (Style B) you can enable or disable contact outputs,

namely, Fail contact and Hold contact. The PH201G (Style B) can output Hold contact

and Fail contact signals. You can set Service Code 53 to “0” for “soft fail” to disable

Fail contact output. When you set Service Code 53 to “1” for “hard fail”, set Service

Code 60 to “2.0” to enable Fail contact output of PH201G (Style B), or set Service

Code 60 to “0.1” to disable Fail contact output of PH201G (Style B).

*ADDR Select address 00 for point communication with 4-20 mA transmission. Addresses 01 to

15 are used in multi-drop configuration (fixed 4 mA output).

T5.3.6E.eps

R

R

(

*

) In case of communication with and without distributor, set "2.0" and "0.1", respectively.

Code Display Function

Function detail

X

Y

Z Ddefault values

Communication

60

*COMM Communication (*)

Set HART communication Off

0

1.0

Set HART communication On

1

On

communication write enable

0

write enable

communication write protect

1

Set communication PH201G*B

2

Without half time check by setting 2.0

0

With half time check by setting 2.1

1

*ADDR

Network address

Set address 00 to 15

00