3. wiring of sensors, 3-1. general precautions, 3-3. installation in: hazardous area-non-incendive – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 35: 3. wiring of sensors -3, 3-1. general precautions -3

Advertising
background image

IM 12D06A03-01E

Installation and wiring 3-3

DISTRIBUTOR

RECORDER

CURRENT OUTPUT

12

180

100

0

Supply

mA-Output

2,5 or 10 m

SENSORS

Safety Barrier

ISC202S only

INPUT

OUTPUT
SUPPLY

COMPUTER

HAND HELD

COMMUNICATOR

MODE

HOLD

FAIL

YES N O

ENT

MEASURE

DISPLA Y
H O L D

MODE

C A L

Figure 3-6. System configuration

3-3. Wiring of sensors
3-3-1. General precautions

Generally, transmission of signals from Inductive Conductivity sensors is at a low voltage and current

level. Thus a lot of care must be taken to avoid interference. Before connecting sensor cables to the

transmitter make sure that the following conditions are met:

• the sensor cables are not mounted in tracks together with high voltage and or power switching cables.

• only standard sensor cables or extension cable are used.

• the transmitter is mounted within the distance of the sensor cables (max. 10 m) + up to 50 m WF10

extension cable.

• the setup is kept flexible for easy insertion and retraction of the sensors in the fitting.

3-3-2. Additional precautions for installations in hazardous areas

Make sure that the total of capacitance and inductances connected to the input terminals of the EXA

ISC202S do not exceed the limits given in the certificate.

This sets a limit to the cable and extensions used.

• The intrinsic safe version of the ISC202S instrument can be mounted in Zone 0 or 1 (ISC202S-B, -N,

-D).

• The sensor can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in the

system certificate is used.

• Ensure that the total of capacitance and inductances connected to the terminals of the EXA ISC202S

do not exceed the limits given in the certificate of the safety barrier or distributor.

• The cable used should preferably have a BLUE colour or marking on the outside.

• Installation for (sensors in Zone 0 or 1):

• Generally, the distributor with input/output isolation has no external earth connection. If there is an

earth connection on the distributor and the external connection of the transmitter is connected to

“protective” earth, the shield of the 2-wire cable may NOT be connected to “protective” earth at the

distributor too.

3-3-3. Installation in: Hazardous Area-Non-Incendive

The EXA ISC202S-N may be installed in a Category 3/ Zone 2/ Div.2 area without the use of safety bar-

riers. Maximum permissible supply voltage 31.5V.

mA

mA

mA

to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in the

range of 6 to 12 mm (0.24 to 0.47 inches).