Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 99

Advertising
background image

IM 12D06A03-01E

Appendix 3-7

2/4

QIS 12D06A03-21E

x

When the message display shows “4 mA,” the output current must be within in the range

shown in Table 1.

x

Press the [ENT] key repeatedly until the message displays shows “12 mA.” When it shows

“12 mA,” the output current must be within the range shown in Table 1.

To skip the current output not needed to be checked, just press the [ENT] key.

x

Press the [ENT] key repeatedly until the message displays shows “20 mA.” When it shows

“20 mA,” the output current must be within the range shown in Table 1.

x

From the above last step, proceed directly to the temperature indication check in Item 3.3.

Table 1

Current Output (Indication)

Output Range

4.0 mA

4 ±0.02 mA DC

12.0 mA

12 ±0.02 mA DC

20.0 mA

20 ±0.02 mA DC

3.4 Temperature

Indication

Check

<Test>

x

From the last step in Item 3.3, Current Output Test, press the [ENT] key repeatedly until the

message display shows “PT1000.”

x

Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-1

and check the temperature readings. Each temperature reading must be within the range.

Table 2-1 (PT1000)

Resistance of

Resistance Box 1

Indication range

1097.3 :

25.0

±0.3°C

1385.0 :

100.0

±0.3°C

x

Press the [ENT] key. The message display will show “30k NTC.”

x

Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-2

and check the temperature readings. Each temperature reading must be within the range.

Table 2-2 (30k NTC)

Resistance of

Resistance Box 1

Indication range

30 k:

25.0

±0.3°C

2.069 k:

100.0

±0.3°C

x

From the above last step, proceed directly to the resistance (conductivity) indication check in

Item 3.5.

3.5 Resistance (Conductivity) Indication Check

<Test>

x

From the last step in Item 3.4, Temperature Indication Check, press the [ENT] key

repeatedly until the message display shows “SEL.10.”

x

Press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by

“RES.1.”

x

Set resistance box 2 to the value shown in Table 3 and check the indication. The resistance

indication must be within the range.