3-6. communication setup, 3-7. general, 3-6. communication setup -19 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual
Page 61: 3-7. general -19

IM 12D06A03-01E
Parameter setting 5-19
5-3-6. Communication setup
Code 60
*COMM.
The settings should be adjusted to suit the communicating device connected
to the output. The communication can be set to HART
®
or to PH 201*B
distribution (see Appendix 2).
*ADDR.
Select address 00 for point to point communication with 4-20mA transmission.
Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).
Code 61
*HOUR
The clock/calendar for the logbook is set for current date and time as reference.
*MINUT
*SECND
*YEAR
*MONTH
*DAY
Code 62
*ERASE
Erase logbook function to clear the recorded data for a fresh start. This may be
desirable when re-commissioning an instrument that has been out of service
for a while.
5-3-7. General
Code 70
*LOAD
The load defaults code allows the instrument to be returned to the default set
up with a single operation. This can be useful when wanting to change from
one application to another.
mA
mA