2-14 Appendix
Sample 4
Concentration of hydrochloric acid

Meas. range

30-40%

Reference temp.

35°C

Temp. compensation

Enter temperature compensation coefficient,

Relate concentration to conductivity by table

START

Service code 20

Service code 31

Service code 35

Select 1

0% output --> 30 %

100% output --> 40%

Service code 55

TEMP

Calibration in Air

Calibration

35 °C (Default)

NaCl --> NO

TC --> YES ( here use [MODE] key to escape )

Calibrate conductivity value in actual operation

e.g. 34% HCl corresponds to conductivity 772 mS/cm

Select 1

Enter conductivity values corresponding to output current values.

Table (below) relates output current to conductivity at 35°C

Service code 21

TC=1.24 (data from customer)

Output

current

(%)

0

20
40
60
80

100

HCl conc.

(%)

30
32
34
36
38
40

Conductivity

(mS/cm)

845
840
772
731
688
644

sample4E.eps