IM 12D06A03-01E

5-18 Parameter setting

5-3-5. User interface

Code 50

*RET.

When Auto return is enabled, the transmitter reverts to the measuring mode

from anywhere in the configuration menus, when no button is pressed during

the set time interval of 10 minutes.

Code 52

*PASS

Passcodes can be set on any or all of the access levels, to restrict access to

the instrument configuration.

Code 53

*Err.05 to Error message configuration. Two different types of failure mode can be set.

*Err.08

Hard fail gives a steady FAIL flag in the display. A fail signal is transmitted on

the mA output when enabled in code 32.

Soft fail gives a flashing FAIL flag in the display. A good example is the dry

sensor for a soft fail.

Code 54

*E5.LS

Limits can be set for shorted and open measurement. E5 (High) is default set

*E6.LIM

to 3 S and must be in the range of 0.10 S to 9.99 S. E6 (Low) is default set to

5 μS and must be in the range of 0.00 μS to 99.9 μS.

Code 55

*%

Linear weight percentage. For some applications the measured parameter

values may be (more or less) linear to concentration. For such applications it

is not needed to enter an output table, but 0 and 100% concentration values

directly can be set.

Code 56

*DISP.

The display resolution is default set to autoranging for conductivity reading. If a

fixed display reading is needed, a choice can be made out of 5 possibilities.

Code

Display

Function

Function detail

X

Y

Z

Default values

User interface

50

*RET.

Auto return

Auto return to measuring mode Off

Auto return to measuring mode On

0

1

1

On

51

Not used

52

*PASS

Passcode

Note # = 0 - 9, where

1=111, 2=333, 3=777

4=888, 5=123, 6=957

7=331, 8=546, 9=847

Maintenance passcode Off

Maintenance passcode On

Commissioning passcode Off

Commissioning passcode On

Service passcode Off

Service passcode On

0

#

0

#

0

#

0.0.0

Off

Off

Off

53

*Err.05

*Err.06

*Err.07

*Err.08

Error setting

Shorted measurement

Soft/Hard

Open measurement

Soft/Hard

Temperature sensor open Soft/Hard

Temp. sensor shorted

Soft/Hard

0/1

0/1

0/1

0/1

1

1

1

1

Hard

Hard

Hard

Hard

54

*E5.L S

*E6.LIM

E5 limit setting

E6 limit setting

Maximum conductivity value

Minimum conductivity value

3

5

S

μS

55

*%

*0%

*100%

Display mA in w/w%

mA-range displayed in w/w% off

mA-range displayed in w/w% on

Set 0% output value in w/w%

Set 100% output value in w/w%

0

1

0

Off

56

*DISP.

Display resolution

Auto ranging display

Display fixed to XXX.X μS/cm

Display fixed to X.XXX mS/cm

Display fixed to XX.XX mS/cm

Display fixed to XXX.X mS/cm

Display fixed to XXXX mS/cm

0

3

4

5

6

7

0

Auto

57-59

Not used