Quality inspection standards, Isc202s 2-wire inductive conductivity transmitter – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 98

Advertising
background image

IM 12D06A03-01E

3-6 Appendix

13-2. ISC202S 2-Wire Inductive Conductivity Transmitter

Quality
Inspection
Standards

ISC202S
2-Wire Inductive Conductivity Transmitter

QIS 12D06A03-21E

1st Edition: Mar. 2007

1. Scope

This inspection standard applies to the ISC202S 2-Wire Inductive Conductivity Transmitter.

2. Inspection

Items

2.1 Insulation resistance test

* 2.2 Dielectric strength test

2.3 Current output test

2.4 Temperature indication check

* 2.5 Resistance (conductivity) indication check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.

3.

Inspection Methods, Standards and Conditions

x

Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least

5 minutes before conducting the tests. For the connections for the insulation resistance test,

follow the instructions in Sections 3.1and for the connections for the dielectric strength test,

follow the instructions in Section 3.2.

x

Use a testing circuit and test equipment shown in Figure 1, or equivalent, for the tests.

x

Performance tests should be done in the inspection mode where the tests from Section 3.3

through Section 3.5 take place in sequence and cannot be retraced. If the reconfirmation of

a test is needed, turn off the power to the transmitter, turn on the power again, and enter the

inspection mode to restart the tests.

3.1 Insulation Resistance Test

x

Apply 500 V DC between the power supply terminals shorted together (+ and –), input

terminals shorted together (11 to 17) and the earth terminal (G). The insulation resistance

must be 100 MΩ or greater.

3.2 Dielectric

strength

test

Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz

or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must

withstand this voltage. (The sensed current should be 10 mA.)

x

Between the power supply terminals shorted together (+ and –) , input terminals shorted

together (11 to 17) and the earth terminal (G)

3.3 Current Output Test

<Preparation>

x

Set resistance box 1 to 30 k:. Wind ten turns of wire onto the ISC40 sensor and set

resistance box 2 to 1 k:.

<Test>

x

Through key operations on the transmitter, enter Service Code 87, input password 070, and

press the [ENT] key.

x

When “ HIF” appears, press the [YES] key.

x

Press the [ENT] key. (The date in day-month-year (last 2 digits) order will appear.)

x

Press the [ENT] key. (The time in hour-minute-second order will appear.)

x

Press the [ENT] key.