Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 32

Advertising
background image

IM 12D06A03-01E

2-20 Specifications

2-14. Control Drawing of ISC202S FF/PB Specification (CSA)

Safe area

Hazardous area

Zone 0 or 1

Safe area

Apparatus

I.S.

interface

I.S.

certified

Terminator

ISC202S-F

or ISC202S-P

I.S.

certified

Terminator

-

+

-

+

Transmitter

Transmitter

Sensor

Connections

Ui = 24 V

or Ui = 17,5 V

Ii = 250 mA Ii = 380 mA

Pi = 1,2 W Pi = 5,32 W

CSA Ex ia Class I, DIV. 1, Group ABCD

T4 for ambient temp. d 55 qC

-

+

-

+

x

Sensor(s) are a thermocouple, RTD's, passive resistive switch devices, or is CSA entity

approved and meet connection requirements.

x

Electrical data of the ISC202S-F & ISC202S-P:
-

Supply and output circuit:

Maximum input voltage Ui = 24 V

Maximum input current Ii = 250 mA

Maximum input power Pi = 1.2 W

Effective internal capacitance Ci = 220 pF;

Effective internal inductance Li = 0 μH.

or

FISCO field device

Maximum input voltage Ui = 17.5 V

Maximum input current Ii = 380 mA

Maximum input power Pi = 5.32 W

Effective internal capacitance Ci = 220 pF;

Effective internal inductance Li = 0 μH.

-

Sensor input circuit:

Maximum output voltage Uo = 14.4V; Maximum output current Io = 20 mA

Maximum allowed external capacitance Co = 600 nF

Maximum allowed external inductance Lo = 88 mH

x

Any CSA approved I.S. interface may be used that meets the following requirements:

Uo d 24 V

Io d 250 mA

Po d 1.2 W

Co t 220 pF + Ccable; Lo t 0 μH + Lcable

or

FISCO field device

Uo d 17.5 V

Io d 380 mA

Po d 5.32 W

Co t 220 pF + Ccable; Lo t 0 μH + Lcable

Installation should be in accordance with Canadian Electrical Code, Part I or CEC, Part I.

Maximum safe area voltage should not exceed 250 Vrms.

x

Electrical data of the ISC202S-B & ISC202S-D (non-incendive):

For Class I, Div.2, Group ABCD the CSA approved I.S. interface is not required, and the

sensor input circuit is non-incendive having the parameters:

Maximum output voltage Uo = 14.4 V; Maximum output current Io = 20 mA

Maximum allowed external capacitance Co = 3.5 μF

Maximum allowed external inductance Lo = 200 mH