Installation and wiring 3-5

3-6. Other sensor systems

To connect other sensor systems, follow the general pattern of the terminal connections as listed below:

11 and 12

: Always used for temperature compensation resistor input.

13 and 17

: Used for the secondary ‘collector’ coil.

15 and 16

: Used for the primary ‘drive’ coil.

14

: Overall screen

Figure 3-8. Terminal identification label

3-6-1. Sensor cable connections using junction box (BA10) and extension cable (WF10)

Where a convenient installation is not possible using the standard cables between sensors and transmit-

ter, a junction box and extension cable may be used. The Yokogawa BA10 junction box and the WF10

extension cable should be used. These items are manufactured to a very high standard and are neces-

sary to ensure that the specifications of the system are not compromised. The total cable length should

not exceed 60 mtr (e.g. 5 m fixed cable and 55 m extension cable).

14 Overall Screen

11

12

12

13

14

14

16

15

13

14

14

16

15

17

11

17

11 Red

12 Blue

15 Core 16 Screen
White Co-axial cable

13 Core 17 Screen
Brown Co-axial Cable

WF10 Cable

TRANSMITTER / CONVERTER

A

C

Overall

shield

B

D

Screen

E

Red

White

Blue

Brown

Thermistor (Temperature sensor)

Ground (Shield)

Secondary Coil

Primary Coil

11
12
17
13
15
16
14

Fig. 3-9. Connection of WF10 extension cable and BA10/BP10 junction box

A-15

B-16

C-13

D-17

E-11

F-12
S-14

temp

> Connections Inductive conductivity