Error messages and explanation, Error messages and explanation -1 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 67

Advertising
background image

IM 12D06A03-01E

Error messages 9-1

9. ERROR MESSAGES AND EXPLANATION

Code Error description

Possible cause

Remedy

E2

Wrong temperature coefficient

Incorrect data entry

See section 5-2-3, 5-2-4, 5-2-5

E3

Calibration out of range (>factor 10)

Wrong unit (mS vs μS)

Defective sensor

Standard error

See section 6-1, 6-2

Replace sensor

Check standard

E4*

Impossible program for Temperature

Compensation

Incorrect data in 5 x 5 Matrix in code

24-28

See section 5-3-3

E5*

Conductivity too high

Incorrect wiring

Defective sensor

Check wiring

Replace sensor

E6*

Conductivity too low

Sensor not submersed

Sensor plugged

Incorrect wiring

Defective sensor

Check installation

Clean sensor

Check wiring

Replace sensor

E7*

Temperature too high

If 30k NTC Temperature < -20° (-4°F)

If Pt 1000 Temperature > 140° (284°F)

Check wiring

Replace sensor

E8*

Temperature too low

If 30k NTC Temperature > 140° (284°F)

If Pt 1000 Temperature < -20° (-4°F)

Check wiring

Replace sensor

E9*

Impossible AIR SET

Too high ZERO

Replace sensor

E10*

EEPROM write failure

Software problem

Unplug the unit, try again

call Yokogawa Service

E15

Impossible adjustment Temperature

Abnormal cable resistance

Check reference see section

5-3-2

E17

Outspan span too small

Temperature compensation span

too small

Max. zero suppression is 90%

Min. temperature span is 50°C

See section 5-2-1

See section 5-2-4, 5-2-5, 5-3-3

E18

Impossible program for Output table

Incorrect data in code 04

See section 5-3-3

E19

Programmed values not accepted

Values exceed preset limits

Try again, read instructions

E20*

DATA LOST

Unauthorized programming

Software problem

Call Yokogawa

E21

Corrupted Eprom

Software failure

Call Yokogawa

mA

mA

mA

mA

*These errors will trigger the FAIL if set to on (default is on.)

NOTE : E6 may occure if the sensor is not submerged in a solution, e.g., no sample is present or

the sensor is left in the air. When E6 is displayed, first check for proper sensor installation.

Advertising