2-7. Control Drawing of ISC202S mA HART® Specification (CSA)

Electrical data of the ISC202S.

- Supply and output circuit (terminals + and -)

Maximum input voltage V

max

= 31.5 V. Maximum input current I

max

= 100 mA.

Maximum input power P

max

= 1.2 W.

Effective internal capacitance C

i

= 22 nF. Effective internal inductance L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 17):

Maximum output voltage V

oc

= 14.4 V. Maximum output current I

sc

= 20 mA.

Maximum allowed external capacitance C

a

= 600 nF.

Maximum allowed external inductance L

a

= 88 mH.

Barriers and power supply should be CSA certified. The specifications must not exceed

the maximum values as shown in the diagram above. Installation should be in

accordance with Canadian Electrical Code, Part I.

Maximum safe area voltage should not exceed 250 V

RMS

.

For Class I, Div. 2, Group ABCD the CSA certified barrier is not required, and the Sensor

input circuit (terminals 11 through 17) is non-incendive having the parameters:

Maximum output voltage V

oc

= 14.4 V.

Maximum output current I

sc

= 20 mA.

Maximum allowed external capacitance C

a

= 3.5μF.

Maximum allowed external inductance L

a

= 200 mH.

The Hand Held Communicator must be of a CSA certified intrinsically safe type in case it

is used on the intrinsically safe circuit in the hazardous area, or of a CSA certified

non-incendive type in case it is used on the non-incendive circuit in the hazardous area.

S afe area

V max = 31.5 V o ltD C
Imax = 100 m A

P max = 1.2 W att

C SA certified
Pow er S upply
(H A R T compatible) )

O utput

Supply

ISC 40S S ensor
terminals 11-17

+

_

G

IS C 40 S S en sor
term in als 1 1 -1 7

H azardous area

Safe area

+

_

G

Functional
earth

Functional
earth

Intrinsically safe design
C SA E x ia C lass I, D iv.1, G roup A B C D ,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

IS C 202S transmitter

+

-

Load

R esistance

Im ax = 100 mA

V max = 31.5 V o ltD C

H azardous area

Functional
earth

+

_

24 volts D C N ominal

Supply V oltage.

C SA certified safety barrier or
pow er supply w ith R int= 300

:

(H A R T compatible)

S uitable values are:

Fo r electrical data:
see text belo w .

Fo r electrical data:
see text belo w .

S uitable values are:

Intrinsically safe design
C S A E x ia C lass I, D iv.1, G roup A B C D ,

T 4 for ambient temp. < 55°C

T 6 for ambient temp. < 40°C

IS C 202S transmitter