5 v. maximum input current i, 100 ma. maximum input power p, 2 w. effective internal capacitance c – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 19: 22 nf. effective internal inductance l, 4 v. maximum output current i, 20 ma. maximum allowed external capacitance c

Advertising
background image

IM 12D06A03-01E

Specifications 2-7

2-3. Control Drawing ISC202S mA HART® Specification (IECEx).

Electrical data of the ISC202S.

- Supply and output circuit (terminals + and -):

Maximum input voltage U

i

= 31.5 V. Maximum input current I

i

= 100 mA.

Maximum input power P

i

= 1.2 W.

Effective internal capacitance

C

i

= 22 nF.

Effective internal inductance

L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 17):

Maximum output voltage U

o

= 14.4 V. Maximum output current I

o

= 20 mA.

Maximum allowed external capacitance

C

o

= 600 nF (for ISC202S-A),

C

o

= 3.5 PF (for ISC202S-N)

Maximum allowed external inductance

L

o

= 88 mH (for ISC202S-A),

L

o

= 200mH (for ISC202S-N)

Barriers and power supply specification must not exceed the maximum values as

shown in the diagram above. These safety descriptions cover most of the commonly

used industry standard barriers, isolators and power supplies.

The Hand Held Communicator must be of a IECEx certified intrinsically safe type in

case it is used on the intrinsically safe circuit in the hazardous area or of a IECEx

certified non-incendive type in case it is used in the non-incendive circuit in the

hazardous area.

ISC 202S ((Inductive C onductivity-transmitter)

Safe area

U o = 31.5 V o lt D C

Io = 100 mA

Po = 1.2 W att

Ex ia or ib C ertified R epeater
Pow er Supply
(HA R T C ompatible)

O utput

Supply

ISC 40S Sensor
term inals 11-17

+

_

G

ISC 40S Sensor
term inals 11-17

Zo ne 0 o r 1

H azardous area

Safe area

+

_

G

Functional
earth

Fanctional
earth

Intrinsically safe design
IEC Ex standard Ex ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

ISC 202S

+

_

Load

R esistance

Ex ia or ib
C ertified safety barrier or pow er
w ith R int=300

:

Io = 100 mA

U o = 31.5 V o lt D C

Zo ne 0 o r 1

H azardous area

Fanctional
earth

+

_

24 volts D C N ominal

Supply V oltage.

Intrinsically safe design

IEC Ex standard Ex ia IIC : T4 for ambient temp. < 55°C

T6 for ambient temp.< 40°C

(HA R T compatible)

(Inductive C onductivity-transmitter)

C ertificate nr. IEC Ex KEM 06.0054X

C ertificate nr. IEC Ex KEM 06.0054X

Advertising