Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 57

Advertising
background image

IM 12D06A03-01E

Parameter setting 5-15

Code Display

Function

Function

detail

X

Y

Z

Default

values

Temperature compensation functions
20

*T.R.°C

Set reference temp.

Use

keys to set value

25 °C

21

*T.C.1

Set temp. coef. 1

Adjust compensation factor

2.10 %

if set to *T.C. in section 5-2-5.

per °C

Set value with

keys

*T.C.2

Set temp. coef. 2

Adjust compensation factor

2.10 %

if set to *T.C. in section 5-2-5.

per °C

Set value with

keys

22

*MATRx

Select matrix

Choose matrix if set to *MATRX

in section 5-2-5, using

keys

H

2

SO

4

, 0 -100ºC, 0.5 - 5%

1

1

H

2

SO

4

H

2

SO

4

, 0 -100ºC, 2.5 - 25%

2

HCl, 0 - 60ºC, 0.5 - 5%

3

HCl, 0 - 60ºC, 1 - 20%

4

HNO

3

, 0 - 80ºC, 0.5 - 5%

5

HNO

3

, 0 -80ºC, 2.5 - 25%

6

NaOH, 0 -100ºC, 0.5 - 5%

7

NaOH, 0 -100ºC, 0.5 - 15%

8

User programmable matrix

9

23

*T1 °C (°F) Set temp. range

Enter 1st (lowest) matrix temp. value

*T2..

Enter 2nd matrix temp. value

*T3..

Enter 3rd matrix temp. value

*T4..

Enter 4th matrix temp. value

*T5..

Enter 5th (highest) matrix temp. value

24

*L1xT1

Enter conductivity

Value for T1

*L1xT2

values for lowest

Value for T2

....

concentration

*L1xT5

Value for T5

25

*L2xT1

Concentration 2

Similar to code 24

26

*L3xT1

Concentration 3

Similar to code 24

27

*L4xT1

Concentration 4

Similar to code 24

28

*L5xT1

Concentration 5

Similar to code 24

29

Not

used