Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 84

Advertising
background image

IM 12D06A03-01E

2-12 Appendix
Sample 2
NaOH concentration

Meas. range

0-10%

Reference temp.

25°C Temp. compensation

Use existing (temperature, concentration) matrix

Service code 55

TEMP

Calibration in Air

Calibration

Service code 22

START

Output Current (%)

0

10
30
60

100

NaOH conc (%)

0
1
3
6

10

Conductivity (mS/cm)

0

53

145
256
359

Enter conductivity values corresponding to output current values.

Existing table (below) relates output current to conductivity at 25°C

Service Code 20

25°C (default)

Service Code 31

Service Code 35

Select 1

0% output --> 0 %

100% output --> 10%

NaCl --> NO

TC --> YES ( here use [MODE] key to escape )

Matrix code 8

Calibrate conductivity value in actual operation

e.g. 3% NaOH corresponds to conductivity 145 mS/cm

(see above table)

sample2E.eps

Select 1