4-2 Operation

ENT

HOLD

FAIL

YES NO

ENT

NO

MODE

YES

MODE

MEASURE
CAL
DISPLAY

HOLD

OUTPUT
SET HOLD
TEMP

SERVICE

Figure 4-1. ISC202 operator interface

4-2. Explanation of operating keys
MODE key

This key toggles between the measuring and maintenance modes. Press once to obtain

access to the maintenance function menu.

CALIB

DISP.1

DISP.2 - (Only when second temp. compensation enabled, section 5.2)

HOLD - (only when enabled, section 5.2)

Press again to return to the measuring mode (press twice when hold is activated).

YES/NO keys

These are used to select choices from the menu.

YES is used to accept a menu selection.

NO is used to reject a selection, or to move ahead to the next option.

DATA ENTRY keys

Is used as a ‘cursor’ key. Each press on this key moves the cursor or flashing digit one

place to the right. This is used to select the digit to be changed when entering numerical

data.

Is used to change the value of a selected digit. Each press on this key increases the value

by one unit. The value can not be decreased, so in order to obtain a lower value, increase

past nine to zero, then increase to the required number.

When the required value has been set using the

and

keys, press

to confirm

the data entry. Please note that the EXA does not register any change of data until the ENT

key is pressed.

key

This is the commissioning mode key. It is used to obtain access to the commissioning

menu. This can only be done with the cover removed or opened. Once this button has

been used to initiate the commissioning menu, follow the prompts and use the other keys

as described above.

Output hold flag

Fail flag

Menu pointer flags

Commissioning

function menu

Commissioning

mode access key

Measure/Maintenance

mode key

Broken line indicates area that can be seen through front

cover

Adjustment keys

> : Choose digit to

adjust

^ : Adjust digit

ENT : Confirm change

Selection keys

YES : Accept setting

NO : Change setting

Key prompt flags

Message display

Main display

Units