Appendix 3-3

3/4

QIS 12D06A03-01E

Table 3 SEL.10

Indication

Setpoint of Resistance Box 2

Indication Range

RES. 1

30.00:

0.300±0.002

RES. 3

800.00:

8.00±0.04

RES. 5

30.00k:

300±2

x

Change the number of turns of wire onto the ISC40GJ sensor from ten to one.

x

Press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by

“RES.6.”

x

Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance

indication must be within the range.

x

Press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by

“RES.7.” There is no need to check here, so just press the [ENT] key.

x

The message display shows “ WAIT ” momentarily, followed by “RES.8.”

x

Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance

indication must be within the range.

x

Press the [ENT] key. The message display shows “ WAIT ” momentarily, followed by

“RES.9.”

x

Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance

indication must be within the range.

x

After the test, press the [ENT] key. The message display shows “READY.”

x

Press the [ENT] key to restart the transmitter.

Table 4 SEL1

Indication

Setpoint of Resistance Box 2

Indication Range

RES. 6

300.0:

300±2

RES. 8

8.000k:

8.00±0.10k

RES. 9

80.00k:

80.0±6.3k