6. control drawing isc202s ma hart® specification, Fm non-incendive design) -10 – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 21

Advertising
background image

IM 12D06A03-01E

Specifications 2-9

Electrical data of the ISC202S.

- Supply circuit (terminals + and -):

Maximum input voltage V

max

= 31.5 V.

Maximum input current I

max

= 100 mA.

Maximum input power P

max

= 1.2 W.

Effective internal capacitance C

i

= 22 nF.

Effective internal inductance L

i

= 35 PH.

- Sensor input circuit (terminals 11 through 17):

Maximum output voltage V

t

= 14.4 V.

Maximum output current I

t

= 20 mA.

Maximum allowed external capacitance

C

a

= 600 nF

Maximum allowed external inductance

L

a

= 88 mH.

If Hand Held Terminal (HHT) is not connected to the power supply lines of the ISC202S (see figure 1):

Any FM Approved barrier or power supply may be used that meets the following requirements.

V

oc

or V

t

d

31.5 V; I

sc

or I

t

d

100 mA; C

a

t

22nF + C

cable

; L

a

t

35PH + L

cable

If HHT is connected to the power supply lines of the ISC202S (see figure 2):

The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of the

HHT and the barrier/power supply to determine the cable parameters.

(V

oc

or V

t

) + V

HHT

d

31.5 V; (I

sc

or I

t

) + I

HHT

d

100 mA;

C

a

t

22nF + C

cable

+ C

HHT

; L

a

t

35PH + L

cable

+ L

HHT

When installing this equipment, follow the manufacturer ’s installation drawing.

Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically Safe

Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70).

Control equipment connected to the barrier/power supply must not use or generate more than 250

Vrms or Vdc.

Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0 Ohm.

In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.

WARNING

- Substitution of components may impair Intrinsic Safety

- To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or

read, understand and adhere to the manufacturer ’s’live maintenance procedures.

ISC40S Sensor
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.

ISC40S Sensor
terminals 11-17
Max. cablelength: 60 mtr.
Cable dia.: 3…12 mm.

Ùnclassified Location

FM Approved

Power Supply

(HART compatible)

Output

Supply

+

_

G

Classified Location

Unclassified Location

+

_

G

Functional
earth

Fanctional
earth

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

ISC202S transmitter

+

-

Load

Resistance

Classified Location

Fanctional
earth

+

_

24 volts DC Nominal

Supply Voltage.

FM Approved safety barrier or
power supply
with Rint = 300

:

(HART compatible)

For electrical data:
see text below.

For electrical data:
see text below.

Intrinsically safe design
FM Class I, Div.1, Group ABCD,

T4 for ambient temp. < 55°C

T6 for ambient temp. < 40°C

ISC202S transmitter

Figure 1

Figure 2

2-5. Control Drawing ISC202S mA HART® Specification (FM Intrinsically safe design)

Application Doc. No.: IKE028-A10 P.5 to P.6