4-1. settings performed in maintenance mode, 4-1-2 selecting items for display – Yokogawa EXA ISC202 2-wire Conductivity Transmitter/Analyzer User Manual

Page 78

Advertising
background image

IM 12D06A03-01E

2-6 Appendix
12-4-1. Settings Performed in Maintenance Mode

12-4-1-1 Calibration with solution of known conductivity

Refer to Sec. 6 for Calibration

1. When is calibration required?

Inductive conductivity meters should be calibrated before they are first used, but should not need peri-

odic calibration after that: the ISC202/ISC40 inductive conductivity measuring system passes a current

through the measured liquid to measure its conductivity; if there is little clearance (less than 30 mm)

around the sensor then this can affect the accuracy of calibration, and the sensor should be calibrated

before use. Also if the sensor surface becomes corroded or contaminated, you should recalibrate it.

2. How is calibration performed

Calibration normally involves measuring a liquid of known conductivity at a known temperature, and

adjusting the meter to read the correct value. Since the reading is affected by the mounting, the sensor

should be calibrated on site after installation.

You can mix standard solutions of KCl or NaCl for calibration. The temperature of the standard solution

should be maintained at the reference temperature. Conductivity tables for NaCl and KCl are shown in

Table 6-1 and 6-2. The other conductivity tables may be found in IEC and other standards. You can

also calibrate the sensor with a solution of arbitrary concentration by calibrating it against a standard

conductivity meter (we recommend Yokogawa's Personal Conductivity Meter). The reference tempera-

ture of both meters should be the same, and as far as possible you should perform the calibration at the

reference temperature, as there will otherwise be temperature compensation errors if the temperature

compensation functions of the two meters are not exactly the same.

When calibration is performed, the cell constant ("Installation Factor") is corrected (see Sec. 5-1-2).

12-4-1-2 Selecting Items for Display

Refer to Sec. 5-1-3 and 5-1-4

1. Selecting Items for Display

You can select and display the following items on the "message" auxiliary display:

• Measured values

• Error messages

• Other messages

The shipping-time default is for temperature to be displayed.

You can change this default to any item on the list in 2. below.

2. What items can be displayed?

Temperature

: Current value

Output

: Current value

Weight%

: Current value

Installation Factor

: See Sec. 6 for Calibration, or Sec. 5-1-2 for Installation Factor

adjustment

Reference Temperature

: See Sec. 5-3-3 for Reference Temperature setting

Temperature Compensation : For default "standard" temperature compensation, NaCl is displayed

When manual temperature compensation is selected, "T.C. coefficient" is displayed.

When matrix temperature compensation is selected, You need to specify "reference temperature", "tem-

perature compensation coefficient", and (non-linear) "compensation characteristic".

3. Examples

You can display output signal value (range 4 to 20 mA) on the "message" auxiliary display to check the

value. For example, if :

Measurement range

: 0 to 100 μS/cm, and

Process value

: 60 μS/cm, then

Output value

: 13.6 mA is displayed.

Advertising