1 gauge 3 wire method, Adjacent 2 gauge method – Yokogawa Standalone MW100 User Manual

Page 102

Advertising
background image

2-14

IM MW100-01E

2.4 Connecting Signal Wires

• 1 Gauge 3 Wire Method

ON

OFF

No.1

No.2

No.3

No.4

No.5

A(+V)

B(L)

C(-V)

D( H)

No.1

ON

No.2

ON

No.3

OFF

No.4

ON

No.5

OFF

Rg

Setting switch

-B12, -B35

R

R

R

Rg

r

r

r

E

e

R: fixed resistance

r: resistance value of lead wire

Rg: resistance value of strain gauge

e: output voltage from bridge

E: voltage applied to bridge

-NDI

Setting

switch

Rg

ON

1 2 3 4 5

1

2

3

4

5

6

7

8

OFF
SW

Bridge head

(701955 or 701956)

SW1

OFF

SW2

ON

SW3

ON

SW4

ON

SW5

OFF

• Adjacent 2 Gauge Method

ON

OFF

No.1

No.2

No.3

No.4

No.5

A(+V)

B(L)

C(-V)

D( H)

No.1

ON

No.2

ON

No.3

OFF

No.4

OFF

No.5

ON

Rg1

Rg2

R

R

E

e

Rg1

Rg2

Rg1

Rg2

Setting switch

-B12, -B35

R: fixed resistance

r: resistance value of lead

wire

Rg: resistance value of strain

gauge

e: output voltage from bridge

E: voltage applied to bridge

-NDI

Setting

switch

Rg1

Rg2

ON

1 2 3 4 5

1

2

3

4

5

6

7

8

OFF
SW

Bridge head

(701955 or 701956)

SW1

OFF

SW2

ON

SW3

ON

SW4

ON

SW5

OFF