Measurement input types, Measurement range, App index – Yokogawa Standalone MW100 User Manual

Page 49

Advertising
background image

1-35

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

1.4 Functions of the 4-CH, High-Speed Universal

Input Module

This module allows up to four inputs of DC voltage, thermocouple, 3-wire RTD, or digital
input (DI) at a minimum measurement interval of 10 ms.

Input terminal (clamp terminal)

Terminal cover

Measurement Input Types

Measurement Input Type

Notation

Do not measure

SKIP

DC voltage

VOLT

Thermocouple

TC

Resistance temperature detector

RTD

DI

DI

Remote RJC

RRJC

Measurement Range

DC Voltage

Measurement Range

Notation

Rated Measurement Range

20 mV

20 mV

–20.000 to 20.000 mV

60 mV

60 mV

–60.00 to 60.00 mV

200 mV

200 mV

–200.00 to 200.00 mV

2 V

2 V

–2.0000 to 2.0000 V

6 V

6 V

–6.000 to 6.000 V

20 V

20 V

–20.000 to 20.000 V

100 V

100 V

–100.00 to 100.00 V

60 mV (high resolution)

60 mVH

0.000 to 60.000 mV

6 V (high resolution)

6 VH

0.0000 to 6.0000 V

1 V

1 V

–1.0000 to 1.0000 V

Thermocouple

Measurement Range

Notation

Rated Measurement Range

Type-R

R

0.0 to 1760.0°C

Type-S

S

0.0 to 1760.0°C

Type-B

B

0.0 to 1820.0°C

Type-K

K

–200.0 to 1370.0°C

Type-E

E

–200.0 to 800.0°C

Type-J

J

–200.0 to 1100.0°C

Type-T

T

–200.0 to 400.0°C

Type-N

N

0.0 to 1300.0°C

Type-W

W

0.0 to 2315.0°C

Type-L

L

–200.0 to 900.0°C

Advertising