App index – Yokogawa Standalone MW100 User Manual

Page 37

Advertising
background image

1-23

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

Stop recording after creating a single file

Single

Fullstop*

Rotate*

Time (s)

Start

File

File

Time (s)

Start Trigger condition met Stop

File

Trigger-wait status

Trigger condition met

Trigger condition met

* The condition for stopping the recording when set to Fullstop and the condition for deleting the file

when set to Rotate is the same as the condition when the recording start action is set to Direct.

Pretrigger Function

When selecting Trigger, a pre-trigger can be set in units of ten percent of the data length.
Writing starts simultaneously with recording, and after a trigger event occurs, the data
remaining after removal of the pretrigger data is written.

Time (s)

Start

Trigger condition met

Trigger point

Stop

File

70% after the trigger

30% before the trigger

* When the pretrigger is set to 30% and

the recording stop action is Single

If the time from when recording is started until the trigger condition is met is shorter than
the pretrigger length, data is written until the data length is reached after the trigger
condition is met.

Trigger point

70% after the trigger

30% before the trigger

Time (s)

Start Trigger condition met Stop

* When the pretrigger is set

to 30% and the recording

stop action is Single

File

1.3 Functions of the Main Module

Advertising