Alarms, Alarms -17, App index alarms – Yokogawa Standalone MW100 User Manual

Page 31

Advertising
background image

1-17

IM MW100-01E

Explanation of Functions

1

2

3

4

5

App

Index

Alarms

The main module compares the measured values against preset alarm values and

outputs alarm signals based on the result from the digital output module. The following

four types of alarms can be output.

Type

Notation Actions

Upper limit alarm

H

Generates an alarm when the measured value

exceeds the alarm value.

Lower limit alarm

L

Generates an alarm when the measured value falls

below the alarm value.

Differential upper limit alarm

DH

Generates an alarm when the difference between

(during differential computation)

the measured values of two alarms exceeds the alarm

value.

Differential lower limit alarm

DL

Generates an alarm when the difference between

(during differential computation)

the measured values of two alarms falls below the alarm

value.

Upper limit on rate-of-change

RH

Generates and alarm when the rate of change in

alarm

rising measured values exceeds the alarm value.

Lower limit on rate-of-change

RL

Generates and alarm when the rate of change in

alarm

falling measured values falls below the alarm value.

Delay upper limit alarm

tH

Generates an alarm when the measured value

remains above the alarm value for the specified time

(delay time).

Delay lower limit alarm

tL

Generates an alarm when the measured value

remains below the alarm value for the specified time

(delay time).

Alarm Value Hysteresis

You can set a width (hysteresis) to the values used to activate and release alarms. Alarm

hysteresis can prevent frequent activation and release of alarms when the measured

value is unstable around the alarm value.

Alarm

setting

Alarm release

Measured vale

Alarm activated

Upper limit alarm

Lower limit alarm

Measured value

Alarm release

Hysteresis

Alarm setting

Alarm activated

Alarm Output Timing

Alarms occur at each measurement interval based on the alarm settings. However when
the measurement interval is 10 or 50 ms, alarms occur at 100 ms intervals based on all
of the data.

Delay Upper Limit Alarm and Delay Lower Limit Alarm

An alarm occurs when the measured value remains below or above the alarm value for
the specified time (delay time). You can set the delay time between 1 and 3600 s for
each channel. Set the delay to an integer multiple of the measurement or MATH interval.

Delay time

Alarm

setting

Alarm release

Delay time

Measured value

Alarm activated

Delay upper limit alarm

Delay lower limit alarm

Measured

value

Alarm cleared

Alarm release

Alarm activated

1.3 Functions of the Main Module

Advertising