CONFIGURATION

103

Real time vs Equivalent time sampling mode

Background

GBS-1000 automatically switches between two
sampling modes, Real-time and Equivalent-time,
according to the number of active channels and the
sampling rate.

Parameter

Real-time
sampling

One-time sampled data is used to
reconstruct a single waveform.
Short-time events might get lost if
the sampling rate gets too high.
This mode is used when the
sampling rate is relatively low.

Equivalent-
time sampling

Multiple numbers of sampled data
are accumulated to reconstruct a
single waveform. Restores greater
waveform details but takes longer
to update the waveform. This mode
is used when the sampling rate
becomes higher.

Real-time /
Equivalent-time
sampling
threshold

Input channel:

Activated

Not activated

Does not matter

1

2

3

4

1G

25G

2.5

250M

500M

Real-time
Sampling

Equivalent-
Time
Sampling

Input Channel

Sa/s