GW Instek GBS-1000 Series User Manual User Manual
Page 173
Advertising
MAINTENANCE
173
3. Press F5 (More) twice.
More
F 5
x2
4. Press F1 (ProbeComp
Menu).
ProbeComp
Menu
F 1
5. Press F1 (Wavetype)
repeatedly to select the
standard square wave.
Wave Type
F 1
6. Press the Auto Set key. The
compensation signal
appears on the display.
Auto Set
7. Press the Display key, then
F1 (Type) twice to select the
vector waveform.
Display
Type
Vectors
F 1
8. Turn the adjustment point on the probe until
the signal edge becomes sharp.
Under
Compensation
Normal
Over