GW Instek GBS-1000 Series User Manual User Manual

Page 173

Advertising
background image

MAINTENANCE

173

3. Press F5 (More) twice.

More

F 5

x2

4. Press F1 (ProbeComp

Menu).

ProbeComp

Menu

F 1

5. Press F1 (Wavetype)

repeatedly to select the
standard square wave.

Wave Type

F 1

6. Press the Auto Set key. The

compensation signal
appears on the display.

Auto Set

7. Press the Display key, then

F1 (Type) twice to select the
vector waveform.

Display

Type

Vectors

F 1

8. Turn the adjustment point on the probe until

the signal edge becomes sharp.

Under

Compensation

Normal

Over

Compensation