First time use – GW Instek GBS-1000 Series User Manual User Manual
Page 24
GBS-1000 Series User Manual
24
First Time Use
Background
This section describes how to connect a signal,
adjust the scale, and compensate the probe. Before
operating GBS-1000 in a new environment, run
these steps to make sure the instrument is
functionally stable and that you are comfortable
operating it.
1. Power On
Follow the procedure on the previous page.
2. Reset system
Reset the system by recalling
the factory setting. Press the
Save/Recall key, then F1
(Default Setup). For factory
setting details, see page49.
Save/Recall
Default
Setup
F 1
2. Connect probe
Connect the probe to Channel1 input terminal and
to the probe compensation signal output (2Vp-p,
1kHz square wave). Set the probe attenuation to
x10.
X10
X1
x1
x10
3. Capture signal
(Auto Set)
Press the Auto Set key. A
square waveform appears in
the center of the display. For
Auto Set details, see page54.
Auto Set