GBS-1000 Series User Manual

126

Trigger coupling

Triggers only on the AC component.

Triggers on AC+DC component.

Frequency
rejection

LF

Puts a high-pass filter and rejects the
frequency below 50kHz.

HF

Puts a low-pass filter and rejects the
frequency above 50kHz.

Noise rejection

Rejects noise signal.

Setup Holdoff and Auto level

Background

The Holdoff function defines the waiting period
before the GBS-1000 starts triggering again after a
trigger point. The Auto level function
automatically adjusts the trigger level to the center
amplitude of the waveform.

Panel operation

1. Press the Trigger menu key

twice.

MENU

MENU

2. To set the Holdoff time,

press F1 (Holdoff) and use
the Variable knob. The
resolution depends on the
horizontal scale.

Holdoff

40.00ns

F 1

Range

40ns~2.5s

Pressing F2 (Set to
Minimum) sets the Holdoff
time to the minimum, 40ns.

Set To

Minimum

F 2

Note

The holdoff function is automatically disabled
when the waveform update mode is in Roll or
Scan mode (page110).