Select probe attenuation level

Background

A signal probe has an attenuation switch to lower
the original DUT signal level to the oscilloscope
input range, if necessary. The probe attenuation
selection adjusts the vertical scale so that the
voltage level on the display reflects the real value
on DUT.

Panel operation

1. Press the Channel key.

CH1

2. Press F4 (Probe) repeatedly

to select the attenuation
level.

Probe

x1

F 4

3. The voltage scale in the

channel indicator changes
accordingly. There is no
change in the waveform
shape.

(x1)

(x10)

(x100)

Range

x1, x10, x100

Note

The attenuation factor adds no influence on the
real signal. It just changes the voltage scale on the
display.