Probe compensation, Ge172), E172) – GW Instek GBS-1000 Series User Manual User Manual

Page 172

Advertising
background image

GBS-1000 Series User Manual

172

8. The calibration for Channel1

starts and ends
automatically, in less than 5
minutes.

Ch1 calibration 1/3

9. When finished, connect the

calibration signal to
Channel2 and press F5.
Channel2 calibration starts.

Done!!

10. Repeat the above step for Channel 3 and 4.

11. When the calibration for all channels is

completed, the display goes back to the default
state.

Probe Compensation

Panel operation

1. Connect the probe between the Channel1 input

and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe attenuation to x10.



X10

X1



x1

x10

2. Press the Utility key.

Utility