8. The calibration for Channel1
starts and ends
automatically, in less than 5
minutes.
Ch1 calibration 1/3
9. When finished, connect the
calibration signal to
Channel2 and press F5.
Channel2 calibration starts.
Done!!
10. Repeat the above step for Channel 3 and 4.
11. When the calibration for all channels is
completed, the display goes back to the default
state.
Probe Compensation
Panel operation
1. Connect the probe between the Channel1 input
and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe attenuation to x10.
X10
X1
x1
x10
2. Press the Utility key.