INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 102

Advertising
background image

5 - 22

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

25 CSID_SUP_INP2P1

Input function Channel 2 #1 Value 0 to 7 (see

section

5.8.4.3 on page 5-27

)

26 CSID_SUP_INP2P2

Input function Channel 2 #2 Value 0 to 7 (see

section

5.8.4.3 on page 5-27

)

27 CSID_SUP_INP2P3

Input function Channel 2 #3 Value 0 to 7 (see

section

5.8.4.3 on page 5-27

)

28 CSID_SUP_INP2P4 Input

function Channel 2 #4 Value 0 to 7 (see

section

5.8.4.3 on page 5-27

)

29 CSID_SYSMODE

Operational mode
sysmode_class

0 - sysmode_simplex1
1 - sysmode_simplex2
2 - sysmode_backup
3 - sysmode_average
4 - sysmode_alloy
5 - sysmode_independent
NOTE:

Not all values are
available.

Available values are
dependent on the value stored
in the CSID_SUP_SNSR
parameter. An invalid value
returns an error.

Invalid values may be made
available by selecting another
value for CSID_SUP_SNSR
than the value currently
programmed.

30 CSID_SYSTESTMODE

System test mode

Simulates a rate based on
source shutter opening.
Is cleared to False/TestOff
every time a new
reset/powerup.

0 - False/TestOff
1 - True/TestOn

31 CSID_SYSASRC

Active source (channel)

Not meaningful in dual
channel modes, such as
Independent and Alloy.

0 - Channel 1
1 - Channel 2

32 CSID_SYSDISP

Display mode (thick/mass)
sysdisp_class

Effects numeric display and
chart plot data.

0 - Thickness
1 - Mass

Table 5-3 E/F command (continued)

#

Command Structure

Comments

Values

Advertising