3 setup parameters, Section, Caution – INFICON STM-2XM 2-Channel Rate/Thickness Monitor User Manual

Page 48

Advertising
background image

3 - 12

PN

07

4-

61

4-

P1

A

STM-2XM Operating Manual

SMDP address. . . . . . . . . . . . . . . . . .Serial communication mode for STM-2XM. If

the SMDP address is 16, STM-2XM uses
RS-232 electrical signaling; otherwise
STM-2XM uses RS-485 signaling.

3.2.1.3 Setup Parameters

CAUTION

Changes in the Setup menu occur immediately. Relays
and analog outputs can activate unexpectedly during
programming.

Sensor setup . . . . . . . . . . . . . . . . . Main parameter that effects the

choices/modes available (see

Table 3-2

).

Tooling, X1:SRC1 . . . . . . . . . . . . . . The ratio between substrate thickness using

Source 1 and Sensor 1 thickness (see

section 7.3, Determining Tooling, on page
7-2

).

Tooling, X2:SRC1 . . . . . . . . . . . . . . The ratio between substrate thickness using

Source 1 and Sensor 2 thickness (see

section 7.3, Determining Tooling, on page
7-2

).

NOTE: In independent chambers, set this

value to zero.

Table 3-2 Sensor setup parameter

Sensor Setup

Available modes

Rate sampling

capable

Backup

capable

1:Single

unshuttered

Simple

No

No

2:Single

shuttered

Simple

Yes

No

3:Dual w/single

shutter

Simple

Yes

Yes

4:Dual

unshuttered

Simple, Averaging, Alloy, Independent

No

If mode is

Simple

5:Dual w/dual

shutters

Simple, Averaging, Alloy, Independent

Yes

If mode is

Simple